Your browser doesn't support javascript.
loading
The research on the comparability of thyroid hormone test results using two different detection systems / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 2953-2954, 2014.
Artigo em Chinês | WPRIM | ID: wpr-460121
ABSTRACT
Objective To analysis of the thyroid hormone test results of Abbott Architect ci16200 automatic biochemical-im-mune analyzer (Abbott system) and Roche Cobas e601 automated immunoassay analyzer (Roche system) .Methods The thyroid hormone(including T4 ,FT4 ,T3 ,FT3 ,TSH ) levels of 93 cases of serum samples were respectively detected by Abbott system (CMIA) and Roche system (ECLIA) ,and the results were analyzed .Results The thyroid hormone test results of the two systems had a good correlation ,and the coefficients of T4 ,FT4 ,T3 ,FT3 ,and TSH were 0 .960 ,0 .962 ,0 .976 ,0 .900 ,and 0 .999(P<0 .01) . However ,there were significant difference of the detection results of the two systems(P<0 .05) .Conclusion The thyroid hormone test results of Abbott system and Roche system are not comparable .

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: International Journal of Laboratory Medicine Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: International Journal of Laboratory Medicine Ano de publicação: 2014 Tipo de documento: Artigo