Your browser doesn't support javascript.
loading
Evaluatoin of the accuracy of 4 impression materials by a 3D optical scanner / 实用口腔医学杂志
Journal of Practical Stomatology ; (6): 27-31, 2015.
Artigo em Chinês | WPRIM | ID: wpr-462135
ABSTRACT

Objective:

To study the 3D accuracy of 4 impression materials.

Methods:

A metal mould with a custom tray was prepared for the measurement.Impression samples of 2 alginate and 2 rubber impression materials were prepared in the metal mould(n =3)re-spectively.The length of three tagged lines,along X,Y,Z axis directions of the samples was measured with a 3D optical scanner.Re-sults120 min after sampling the variable rates of the length along X,Y,Z axis of Zhermark alginate samples were 3.615%,3.037%and 2.836%,those of CAVEX alginate samples were 2.836%,3.358% and 4.276% respectively;8 h after sampling the variable rates of Impregum Penta polyether rubber samples were 0.039%,0.071% and 0.057%,those of ExpressXT Penta H addition-curing silicone rubber samples were 0.033%,0.088% and 0.084% respectively.

Conclusion:

The variable rates of the 4 impression materi-als are in accordance with ISO1563∶1990.The variable rate of alginate impression material is greater than that of rubber.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Practical Stomatology Ano de publicação: 2015 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Practical Stomatology Ano de publicação: 2015 Tipo de documento: Artigo