Performance evaluation of Sysmex XN-9000 automated hematology analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 1373-1375, 2015.
Artigo em Chinês | WPRIM | ID: wpr-463311
ABSTRACT
Objective To evaluate the performance of Sysmex XN‐9000 automated hematology analyzer .Methods According to international and domestic standards ,performance of analyzer was evaluated .Results The within‐batch and between‐batch preci‐sion ,carryover pollution rate ,linearity range and the accuracy of Sysmex XN‐9000 analyzer were all conform to related require‐ments .Leukocyte classification results compared with manual classification ,the correlation of neutrophil ,lymphocyte ,monocyte and eosinophil were fine ,but correlation of basophil was not very ideal .Conclusion The performance of Sysmex XN‐9000 analyzer could be satisfying ,could meet the needs of clinical inspection and diagnosis and treatment .

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Guia de Prática Clínica Idioma: Chinês Revista: International Journal of Laboratory Medicine Ano de publicação: 2015 Tipo de documento: Artigo