Your browser doesn't support javascript.
loading
Determination of dichloromethane in drinking water by purge and trap gas chromatography-mass spectrometry / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 1672-1673, 2015.
Artigo em Chinês | WPRIM | ID: wpr-467938
ABSTRACT
Objective To establish a method for determination of dichloromethane in drinking water by purge and trap gas chro‐matography‐mass spectrometry .Methods Dichloromethane in drinking water was purged and trapped ,then desorbed and detected by using gas chromatographymass spectrometry .Quantitive analysis was conducted by using standard curve under selected ion mo‐nitoring (SIM) mode .Results The operation of this method was simple ,and its limit of detection (LOD) was 0 .005 ng/mL .The recovery rate was higher than 97% ,and the relative standard deviation was less than 5% .Conclusion The method may be appro‐priate to measure dichloromethane in drinking water .

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: International Journal of Laboratory Medicine Ano de publicação: 2015 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: International Journal of Laboratory Medicine Ano de publicação: 2015 Tipo de documento: Artigo