Your browser doesn't support javascript.
loading
Investigate the accuracy of real-time continuous glucose monitoring system in the type 2 diabetes subjects / 天津医药
Tianjin Medical Journal ; (12): 936-938, 2015.
Artigo em Chinês | WPRIM | ID: wpr-477966
ABSTRACT
Objective To investigate the accuracy of real-time continuous glucose monitoring(RT-CGM)system in the type 2 diabetes mellitus(T2DM). Methods A total of 318 subjects hospitalized between 2013 May to 2014 August were recruited. Each subject received a RT-CGM system. In order to calibrate the interstitial glucose level, finger-stick blood glucose were measured at least four times every day. Patients were divided into three groups according to their finger-stick blood glucose level≤6.99 mmol/L, 7.0-10.0 mmol/L and≥10.01 mmol/L. Pearson correlation was used to analyze the relationship between the finger-stick glucose and interstitial glucose gained from RT-CGM. Clarke error grid analysis was used to analyze the accuracy of RT-CGM. Results (1)A total of 2, 815 glucose meter values from finger-stick were com?paired with glucose from RT-CGM. A positive relationship(r=0.847)were found between CGM and finger-stick blood glu?cose. The correlation was closer with higher finger-stick blood glucose levels(r=0.457, 0.648 and 0.852 respectively in≤6.99 mmol/L group, 7.0-10.0 mmol/L group and≥10.01 mmol/L group. P0.05). Conclusion RT-CGM provides accu?rate blood glucose values to guide the treatment of diabetes, and the accuracy is more reliable in high glucose environment.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Tianjin Medical Journal Ano de publicação: 2015 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Tianjin Medical Journal Ano de publicação: 2015 Tipo de documento: Artigo