Preparation and Application of Internal Quality Control Substance of Free Hemoglobin / 现代检验医学杂志
Journal of Modern Laboratory Medicine
; (4): 117-119, 2015.
Article
em Zh
| WPRIM
| ID: wpr-482582
Biblioteca responsável:
WPRO
ABSTRACT
Objective To prepare an internal quality control substance of free hemoglobin (FHb),and evaluate its perform-ance.Methods ① 5 normal whole human blood samples with known hemoglobin (Hb)concentration was selected and dilu-ted with distilled water in the ratio of 1∶500,1∶1 000,1∶2 000,1∶4 000 respectively.After 20 FHb solutions was fully mixed,the concentration of FHb solutions was tested and compared with its theoretical data.②One of the highest value and the lowest value solutions were chosen as qualitye control substance,packed and stored in -20℃ fridge.Each group was an-alyzed 20 days continuously in the routine conditions,where average (-x )of the groups,standards deviation (s )and coeffi-cients of variation (CV)were calculated and Levey-Jennings control chart was conducted.Then,the test was repeated every week,and the data on the chart was updated accordingly until (CV)of the reagents were changed.The average (-x)of the-groups,standards deviation (s)and coefficients of variation uality control material was calculated during the whole test peri-od.Results ①The theoretical value (-x ±s)of the concentration of free hemoglobin was 125.44±93.04 mg/L,and the actu-al value was 125.22±93.08 mg/L,there was no statistic significance (t value was 0.706,P >0.05).② During 20 continu-ously daystest,the detected values (-x ±s)of the highest quality control material was 303.55±3.70 mg/L;the values of the lowest quality control material was 69.29 ± 1.88 mg/L.The coefficients were 1.22% and 2.68% respectively.Both the highest and the lowest quality control material were under control during the whole detecting period.The (-x ±s )values of the highest was 302.56±3.99 mg/L,the CV value of it was 1.32%;The (-x ±s)values of the lowest was 69.04±1.88 mg/L,the CV value of it was 2.71%.Conclusion The preparation method of internal quality control substance of free hemoglo-bin is convenient and stable,and was applicable in blood station’s and clinical laboratory.
Texto completo:
1
Índice:
WPRIM
Tipo de estudo:
Guideline
Idioma:
Zh
Revista:
Journal of Modern Laboratory Medicine
Ano de publicação:
2015
Tipo de documento:
Article