Quantitative evaluation of the measurement accuracy of 2 three-dimensional facial scanners / 实用口腔医学杂志
Journal of Practical Stomatology
;
(6): 37-42, 2016.
Artigo
em Chinês
| WPRIM
| ID: wpr-486005
ABSTRACT
Objective:
To evaluate the actual measurement accuracy of 2 three-dimensional(3D)facial scanners for real person.Methods:
3D digital face models of 1 0 volunteers with normal ficial form were obtained by 3dMD and FaceScan facial scanners respec-tively.The measurement values of 1 0 feature lengths and 5 feature angles were measured on each 3D model by the software respective-ly.The reference values of all characteristics were acquired by line laser scanner (Faro)with high accuracy.Statistical and surveying analysis were taken between the measurement values and reference values.Facial morphology measurement error and actual accuracy of facial scanners were obtained finally.Data were statistically analysed.Results:
The length measurement accuracy of 3dMD and FaceS-can was(-0.37 ±0.68)mm and (-0.29 ±0.53)mm(P =0.223),the angle measurement accuracy was (-0.22 ±2.1 4)°and (0.1 2 ±2.69)°(P =0.428),respectively.Conclusion:
The 3D data of ficial morphology obtained by the 2 scanners are not signifi-cantly different.