Comparison of Different Assay System for C-reactive Protein Detection / 现代检验医学杂志
Journal of Modern Laboratory Medicine
;
(4): 127-130, 2016.
Artigo
em Chinês
| WPRIM
| ID: wpr-493762
ABSTRACT
Objective To evaluate comparability of two different assay system for detecting CRP.Methods Following the profile of Clinical and Laboratory Standard Institute (CLSI)document EP9-A2,50 blood samples with anti-coagulant ED-TA-2K were collected from emergency patients at Changhai Hospital.The test result of samples by the i-CHROMA Reader was compared and evaluated with those by Beckman Immage 800.Results The linear regression equation for plasma CRP wasY=1.076 5X-3.031 5,R2=0.986.The linear regression equation for whole blood CRP wasY=0.882 6X-1.180 8, R2=0.931 1.For whole blood samples with low HCT (<30.45%).Used correction equationCRP (after corrected)=CRP (before corrected)/(1-HCT).The regression equation (after corrected)wasY=1.006 8X-3.612 2,R2=0.950 9.Con-clusion CRP concentration detected by i-CHROMA showed good correlation and comparability compared to laboratory ref-erence system by using plasma samples.Results form whole blood samples with low HCT should be corrected to improve comparability.