Determination of Cefotiam in Human Plasma Using Liquid Chromatography-Tandem Mass Spectrometry / 分析化学
Chinese Journal of Analytical Chemistry
;
(12): 876-881, 2016.
Artigo
em Chinês
| WPRIM
| ID: wpr-494384
ABSTRACT
In order to evaluate the pharmacokinetic profile of cefotiam hexetil hydrochloride tablet in Chinese healthy volunteers, a sensitive, specific and rapidprotein precipitation-liquid chromatography-tandem mass spectrometry method was developed and validated in human plasma. Chromatographic separation was achieved on a Waters Symmtry-C18 column (50 mm × 4. 6 mm, 5 μm), using a gradient mobile phase consisting of methanol and 1 mmol/ L ammonium acetate in water at a flow rate of 1. 0 mL/ min. Cefotiam and diazepam (internal standard) were detected without interference in the multiple reaction monitoring (MRM) mode with positive electrospray ionization. The calibration curve was linear from 5. 0 ng / mL to 5000 ng / mL (r>0. 99) with limit of quantitation of 5. 0 ng / mL. The assay met the published acceptance criteria. This rapid, sensitive and reproducible method was successfully applied to the pharmacokinetic study of cefotiam hexetil hydrochloride tablet in healthy Chinese volunteers and therefore provided a considerable mirror for quantification of other cephalosporins in human matrix.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Idioma:
Chinês
Revista:
Chinese Journal of Analytical Chemistry
Ano de publicação:
2016
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS