Research on quality testing for active spot scanning proton and heavy ion accelerator / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection
;
(12): 621-625, 2016.
Artigo
em Chinês
| WPRIM
| ID: wpr-496842
ABSTRACT
Objective To study the quality testing of dose delivery system of the active spot scanning proton and heavy ion accelerator,in order to provide the reference for the quality control of related equipment.Methods In the four therapy rooms,both 0.6 cc chambers and Gafchromic EBT3 films were used,respectively,to test the accelerator for dose reproducibility,dose linearity,dose stability,depth dose distribution,beam scanning position deviation and radiation field uniformity in each therapy room.Results Dose reproducibility variation coefficients are all less than 1.5%,dose linearity's maximum deviations less than 2%,dose stability's deviations less than 2%,depth dose distribution stability within 2%,beam scanning position deviation less than 1 mm,consistency of irradiation field's deviation less than 2 mm,and flatness within ± 5%.Conclusions The indicators about quality testing for the active spot scanning proton and heavy ion accelerator are all in line with the requirements of IEC standards draft.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Tipo de estudo:
Guia de Prática Clínica
Idioma:
Chinês
Revista:
Chinese Journal of Radiological Medicine and Protection
Ano de publicação:
2016
Tipo de documento:
Artigo