Objective:
To develop a method for the determination of palladium in bendamustine hydrochloride by GFAAS. Meth-odsThe sample was destroyed by heat, and the content of palladium was determined by GFAAS with the detection wavelength of 247. 6 nm.Results:
The absorbance and the content of palladium showed a good linear relationship within the range of 20-60 ng· ml-1(r=0. 998 4). The average recovery of palladium was 102. 9%(RSD=1. 7%, n=9).Conclusion:
The method is sensitive and simple, which can be used for the determination of palladium in bendamustine hydrochloride.