Your browser doesn't support javascript.
loading
Uncertainty Evaluation for the Determination of Pseudo-ginsenoside GQ in Human Plasma by HPLC-MS/MS / 医药导报
Herald of Medicine ; (12): 262-267, 2017.
Artigo em Chinês | WPRIM | ID: wpr-511196
ABSTRACT
Objective To evaluate the uncertainty of the pseudo-ginsenoside GQ (PGQ) concentration in human plasma by HPLC-MS/MS.Methods The whole process of PGQ determination by HPLC-MS/MS in human plasma was evaluated and the uncertainty caused by repeatability,weighing,standard solution preparation,biological sample preparation,extraction recovery process,recovery,instrument precision and calibration curve fitting were evaluated,respectively.The combined and expanded uncertainty values were both calculated.Results The expanded uncertainty values for low (15.16 ng·mL-1),medium (2 516.67 ng·mL-1) and high (3 902.00 ng·mL-1) levels of PGQ were 1.39,177.74 and 262.69 ng·mL-1,respectively (P =95 %,k =2).Conclusion The uncertainty of the PGQ determination in human plasma by HPLC-MS/MS is mainly caused by recovery,repeatabihty and sample preparation at low concentration,by sample preparation and recovery at medium and high concentration.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Herald of Medicine Ano de publicação: 2017 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Herald of Medicine Ano de publicação: 2017 Tipo de documento: Artigo