Your browser doesn't support javascript.
loading
Comparison and evaluation of abnormal result detection methods of platelet / 医疗卫生装备
Chinese Medical Equipment Journal ; (6): 97-100, 2017.
Artigo em Chinês | WPRIM | ID: wpr-511344
ABSTRACT
Objective To compare the clinical values of impedance method,optical method and microscopy when used to detect platelet abnormal results.Methods Platelet re-examinations by optical method and microscopy were carried out in case of low confidence degree in platelet test with impedance method by XE-2100 automatic hematology analyzer,and then the results by the three methods were compared.Results Most of the low-confidence-degree results by impedance method could be corrected by re-examination by optical method,and the remained had to turn to microscopy due to unsatisfied requirements of the instrument.Conclusion Optical method has to be involved to correct the platelet abnormal results by XE2100 automatic hematology analyzer,and microscopy should be applied in case optical method doesn't work.The three methods gains advantages and disadvantages of themselves,and can be supplementaries for one another.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: Chinese Medical Equipment Journal Ano de publicação: 2017 Tipo de documento: Artigo