Objective: To evaluate the noise characteristic of different reconstruction type CT image by using the noise power spectrum (NPS), and analyze the correlation between this results and subjective vision noise evaluation. Methods: QA water phantom of CT equipment was scanned and 8 common reconstruction algorithms were applied to dispose image. NPS peak value, the peak frequency and standard deviation (SD) were compared with the subjective evaluation measurements, such as granularity, contrast, sharpness and optical noise level, by using the correlation analysis. Results: Each reconstruction algorithm owned different peak value and peak frequency. Granularity negatively correlated with the peak frequency. A positive correlation was found between contrast and peak value. Sharpness was positively correlated with both peak value and peak frequency. All of the subjective evaluation measurements were correlated with SD. Conclusion: Comparing with SD, NPS can reflect both intensity and morphological feature of the noise and possess applicative potential as a more comprehensive evaluation index.
Palavras-chave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: China Medical Equipment Ano de publicação: 2017 Tipo de documento: Article