Development and validation of a stability-indicating RP–HPLC method for estimation of atazanavir sulfate in bulk / 药物分析学报
Journal of Pharmaceutical Analysis
; (6): 134-140, 2017.
Article
em Zh
| WPRIM
| ID: wpr-513419
Biblioteca responsável:
WPRO
ABSTRACT
A stability-indicating reverse phase–high performance liquid chromatography (RP–HPLC) method was developed and validated for the determination of atazanavir sulfate in tablet dosage forms using C18 column Phenomenix (250 mm×4.6 mm, 5μm) with a mobile phase consisting of 900 mL of HPLC grade methanol and 100 mL of water of HPLC grade. The pH was adjusted to 3.55 with acetic acid. The mobile phase was sonicated for 10 min and filtered through a 0.45μm membrane filter at a flow rate of 0.5 mL/min. The detection was carried out at 249 nm and retention time of atazanavir sulfate was found to be 8.323 min. Linearity was observed from 10 to 90μg/mL (coefficient of determination R2 was 0.999) with equation, y=23.427x+37.732. Atazanavir sulfate was subjected to stress conditions including acidic, alkaline, oxidation, photolysis and thermal degradation, and the results showed that it was more sensitive towards acidic degradation. The method was validated as per ICH guidelines.
Texto completo:
1
Índice:
WPRIM
Idioma:
Zh
Revista:
Journal of Pharmaceutical Analysis
Ano de publicação:
2017
Tipo de documento:
Article