Your browser doesn't support javascript.
loading
The influence of different inclination of cusp on the stress of an implant supported denture / 实用口腔医学杂志
Journal of Practical Stomatology ; (6)1996.
Artigo em Chinês | WPRIM | ID: wpr-539355
ABSTRACT

Objective:

To analyze the influence of different inclination of cusp on the stress of an implant supported denture.

Methods:

A spiral CT was used to scan a corpus' head.The model of mandible,teeth and implant was established by ANSYS 6.1 software. The superstructure of the denture had three different inclination of cusp.The stress distribution of the models under loading was analyzed.

Results:

When bearing non central loading, the maximum equivalent stress (max EQV) of the abutment was situated at the first screw of the abutment.When bearing central loading, the maximum equivalent stress(max EQV) of the abutment was situated at the cervix. Under all loading, the max EQV of the implant was situated at the cervix, the max EQV of mandible was situated at the upper cortical layer around the cervix of the implant. Higher inclination of cusp and non central loading lead to higher stress.

Conclusion:

The location of the max EQV of the abutment is consistent with the location of the abutment fracture; lower inclination of cusp and flatsurface teeth are benefit to the stress distribution of the implantation.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Practical Stomatology Ano de publicação: 1996 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Practical Stomatology Ano de publicação: 1996 Tipo de documento: Artigo