Optimal Beam Tube Potential for the CsI/a:Si Flat Panel Digital Chest Radiography:Whether High KV Technique Still Suitable for Digital Equipment / 实用放射学杂志
Journal of Practical Radiology ; (12)1996.
Artigo em Chinês | WPRIM | ID: wpr-542957
ABSTRACT
Objective To investigate the optimal beam tube potential for amorphous silicon/cesium iodide flat panel digital chest radiography(indirect DR). Methods(1)60 healthy adults were undergone digital posteroanterior(PA) chest radiography at 80,100,150 kV respectively.The images quality of these three were compared by 5 independent observers.Several normal anatomy parts of the chest and the overall impression of the images were evaluated;(2)PA chest exposures of the anthropomorphic chest phantom to produce a constant ESDs(entrance skin doses) and exit dose were made using tube potential at 80,100,120,150 kV.The simulated chest nodules were scored. Results The quality of the adult chest radiography and the manifestation of the simulated chest nodules were superior at the lower tube voltages.The contrast of the images were best at 80 kV,while the appearance of the images look like that of the low potential radiography. Conclusion The optimal tube potential for chest DR is considered to be 80~100 kVp,while the high KV technique is no longer suitable for digital equipment.