Your browser doesn't support javascript.
loading
Determination of Selenium in Water Sample by Hydride Generation-Atomic Fluorescence Spectrometry with Microwave Digestion / 环境与健康杂志
Journal of Environment and Health ; (12)1989.
Artigo em Chinês | WPRIM | ID: wpr-545084
ABSTRACT
Objective To establish a method for detection of selenium in water sample. Methods The method of digestion with nitric acid-hydrogen peroxide was used. Results The calibration curve for selenium was 0-100 ?g/L, r=0.999 2, detection limit 0.42 ?g/L and the recovery of standard addition 98.9% to 102.3%. Conclusion This method is simple, rapid, sensitive and of accurate, the method has been applied to the sample analysis with satisfactory results.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Environment and Health Ano de publicação: 1989 Tipo de documento: Artigo