A GC method is described for measuring fentanyl in man plasma. Fentanyl was extracted at basic pH with cyclohexane-isobutyl alcohol (1973) with Ro21-2212 added as internal standard and fentanyl derivative as earner. The drug was back-extracted in H2SO4, then the extract was made basic and recxtracted with ethyl ether-dichloromethane (91). The organic phase was evaporated at 40℃ under N2 and the residue was dissolved in ethanol. A. portion (2?l) was analyzed on a wide-bore capillary column (10 m ? 0.53 mm) of HP-1 (2.65 ?m), operated at 255℃ with N2 as carrier gas (8 ml/min) and N -P detector. The calibration graph was linear within the ranges of 0.25-100 ng/ml with a corelation coefficient of 0.9996 and the detected limit was 0.2 ng/ml. Mean recovery was 99.02% ? 6.81%. The coefficient of variation for the within- and between-run were all less than 8%. No interference was found from endogenous compounds, metabolites of parent drug or other commonly used drugs. The method was applied to therapeutic drug monitoring and pharmacokinetic studies.