Detection and Handling of Carry-over of Automatic Biochemistry Analyzer / 医学研究杂志
Journal of Medical Research ; (12)2006.
Artigo em Chinês | WPRIM | ID: wpr-561566
ABSTRACT
Objective To investigate the detection and treatment of carry-over of automatic biochemistry analyzer. Methods We detected the sample carry-over by determining one high level quality control serum and three zero levels serum continuously in five days. We used mixed patients serum to detect the carry-over come from reagent probe, muddler and cuvette by arranging test sequence of different item, combination and the times of rerun. We also used different methods such as immersion, cleaning probe, extended priming to handle the reagent carry-over.Results The sample carry-over was -0.37%~0.48% on new analyzer and 0.51~1.17% on old analyzer, respectively. The carry-over of CHO, TG, HDL on reagent probe and muddler of TBA was 86.9%~208.7% and 86.0%~256.5% on old analyzer, respectively. There was no reagent carry-over among other items on old analyzer. There was no carry-over on reagent probe on new analyzer, the carry-over of different items on cuvette of TBA was -0.9%~1.7%. After immersion, cleaning and extended priming, the carry-over of CHD, TG, HDL on TBA reagent probe fell to -1.1%~2.7% and -0.07%~0.09%, respectively.Conclusions The methods which have been mentioned above can detect the sample and reagent carry-over. The carry-over can be reduced by immersion, cleaning, extended priming and testing in different unit.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: Journal of Medical Research Ano de publicação: 2006 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: Journal of Medical Research Ano de publicação: 2006 Tipo de documento: Artigo