Establishing the method of measurement of glucose concentration in brain microdialysate by HPLC-NCSE100-ED / 中国药理学通报
Chinese Pharmacological Bulletin
; (12)1986.
Article
em Zh
| WPRIM
| ID: wpr-561684
Biblioteca responsável:
WPRO
ABSTRACT
Aim To establish the rapidly detecting method of glucose concentration in brain microdialysate.Method Using high-performance liquid chromatography (HPLC) with electrochemical detector (ED) and Nano-Copper screen-printed electrode (NCSE100)-Flow injection analysis the glucose concentration in brain microdialysate were detected.Result The related coefficient (r) of glucose standard curve was 0.9995. The variation coefficient (CV) within a day was 4%~7%. The striatum extracellular glucose concentration in rats was(2.4?0.8) mmol?L-1. These results indicated that our method was reliable.Conclusion Using HPLC with ED, NCSE100, and flow injection analysis the glucose concentration in brain microdialysate were detected, which has certain applied worth.
Texto completo:
1
Índice:
WPRIM
Idioma:
Zh
Revista:
Chinese Pharmacological Bulletin
Ano de publicação:
1986
Tipo de documento:
Article