The adsorption of inspired desflurane due to incorrectly mounted air dryer filter of the heated humidifier: A case report
Anesthesia and Pain Medicine ; : 141-144, 2013.
Artigo em Coreano | WPRIM | ID: wpr-56832
ABSTRACT
Heated humidifiers are used to prevent respiratory complications associated with mechanical ventilation with an endotracheal tube. One of the heated humidifiers uses an air dryer filter in order to remove humidity from air before it enters the anesthetic machine. We report a case in which the air dryer filter was unintentionally misconnected to the inspiratory limb of the circuit, which caused insufficient volatile anesthetic concentration during anesthesia induction.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Respiração Artificial / Adsorção / Extremidades / Sílica Gel / Temperatura Alta / Umidade / Isoflurano / Anestesia Idioma: Coreano Revista: Anesthesia and Pain Medicine Ano de publicação: 2013 Tipo de documento: Artigo