Measurements of Cardiac Output of a New Bileaflet Mechanical Heart Valve Prosthesis Made in China / 航天医学与医学工程
Space Medicine & Medical Engineering ; (6)2006.
Artigo em Chinês | WPRIM | ID: wpr-580813
ABSTRACT
Objective To observe the cardiac output (CO) of animals and patients underwent valve replacement with Jiuling bileaflet mechanical valve prosthesis.Methods 1) animal experimentsCOs of six sheep that underwent mitral valve replacement with 21 mm prosthesis were measured with open cardiac catheterization during operation.Echocardiography and open cardiac catheterization under dobutamine stress were performed in two sheep 30 months post operation.2) patient measurementsCOs of 14 cases of aortic valve and 10 cases of mitral valve replacements were measured with open cardiac catheterization,and after 12 months,they were measured with echocardiography.Results 1)animal experimentsthe mean CO of 6 sheep was 2.5 L/min during operation with catheterization,and it of two sheep 30 months post-implant was 3.0 L/min with echocardiography and 2.9 L/min with catheterization.2)patient measurementsthe mean cardiac index of four patients with 21 mm valve replacement was (2.55?0.27) L/(min?m2) with means of catheterization,and was (2.84?0.13) L/(min?m2) with echocardiography after 12 months.Conclusion It is demonstrated that cardiac function of animals and patients return to normal after they underwent the valve replacement with new China-made heart valve prosthesis (Juling).The new heart valve prosthesis has excellent hemodynamic performance.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Space Medicine & Medical Engineering Ano de publicação: 2006 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Space Medicine & Medical Engineering Ano de publicação: 2006 Tipo de documento: Artigo