Research on Comparison of Imaging Parameters in Intravascular Ultrasound Elastography / 航天医学与医学工程
Space Medicine & Medical Engineering ; (6)2006.
Artigo em Chinês | WPRIM | ID: wpr-580814
ABSTRACT
Objective To make the validation of intravascular ultrasound elastography (IVUSE) for detecting the plaque.Methods A plain strain model of vascular plaque was established and then the RF signals of IVUS were produced.The radial displacement distribution and strain distribution were calculated through 1D numerical optimization based on the RF signals.Then the elastic moduli distribution was reconstructed by finite element method (FEM) and iteration method.Results The three elastic parameters,displacement/strain/moduli,all could detect the plaque.Elastic moduli presented the information of the shape and size of the plaque,meanwhile,the displacement and strain distributions could not give the information clearly.Conclusion Elastic modulus is proved to be superior to the other two parameters in detecting the plaque,whereas it is subject to the accuracy of the strain's estimation.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Space Medicine & Medical Engineering Ano de publicação: 2006 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Space Medicine & Medical Engineering Ano de publicação: 2006 Tipo de documento: Artigo