Research of effect of stray capacitance on measurement accuracy in tissue impedance spectroscopy measurement / 医疗卫生装备
Chinese Medical Equipment Journal ; (6)1993.
Artigo em Chinês | WPRIM | ID: wpr-583779
ABSTRACT
In the tissue impedance spectroscopy measurements ,four-electrode method are normally used. A measurement error caused by stray capacitance between lines of drive electrode and receive electrode is not a trivial problem. This paper studied this error and reached the results This stray capacitance introduces errors which increase with frequency and has more effect on the imaginary part than on the real part in the measurement frequency range.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Medical Equipment Journal Ano de publicação: 1993 Tipo de documento: Artigo