Your browser doesn't support javascript.
loading
Application of Kalman filter to measurement of sound power for medical ultrasonic diagnostic equipment / 医疗卫生装备
Chinese Medical Equipment Journal ; (6)1989.
Artigo em Chinês | WPRIM | ID: wpr-586116
ABSTRACT
A data processing method for the sound power of the medical ultrasonic diagnostic equipment is introduced in this paper. The sound power for the transducer of a medical ultrasonic diagnostic equipment is measured for many times, and then the mathematical model based on the measurement data above is processed with Kalman filter.The output fluctuations for sound power of the equipment and its transducer as well as the measurement errors of sound power meter are eliminated to achieve the true value of sound power. Compared with other mean methods, the method above can improve the dicreteness of the measurements, and thus is worth being applied to processing other measured data.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: Chinese Medical Equipment Journal Ano de publicação: 1989 Tipo de documento: Artigo