Automated capillary zone electrophoresis for analysis of human serum protein in clinical application / 中华检验医学杂志
Chinese Journal of Laboratory Medicine
; (12)2001.
Article
em Zh
| WPRIM
| ID: wpr-586612
Biblioteca responsável:
WPRO
ABSTRACT
Objective The automated capillary zone electrophoresis (CE) is used for human serum protein analysis and compared with the results of agarose gel electrophoresis (AGE) for human serum protein analysis.Discuss its′ practical application in clinical.Methods With the Capillarys ?1-?2+reagent set, proteins in serum were separated at 7 kV for 4 min in 15.5 cm?25.0 ?m fused-silica capillaries (n=8) at 35.5℃ in a pH 10 buffer with online detection 200 nm.Results Capillary zone electrophoresis compared with agarose gel electrophoresis(AGE;Hydrasys-Hyry; Hydragel protein(e)15/30/54 reagent set;Sebia)in precision, relatives, linearity and potential interferences were analyzed.A samples without Para-protein (n=165) were performed relation test. The coefficient of relation between CE and AGE is 0.929 for albumin, 0.924 for for ?1-globulin, 0.841 for ?2-globulin, 0.789 for ?-globulin, 0.926 for ?-globulin. There was no significant difference (P
Texto completo:
1
Índice:
WPRIM
Idioma:
Zh
Revista:
Chinese Journal of Laboratory Medicine
Ano de publicação:
2001
Tipo de documento:
Article