Evaluation on adequacy of hemodialysis by on-line clearance monitoring and Kt/V of Correlation Analysis / 中国医师杂志
Journal of Chinese Physician ; (12): 177-178,181, 2012.
Artigo em Chinês | WPRIM | ID: wpr-589785
ABSTRACT
ObjectiveTo evaluate the adequacy of hemodialysis by online clearance monitoring (OCM) and Kt/V of correlation analysis.MethodsThe Kt/V values of 48 maintenance hemodialysis patients in the Department of Nephrology of General Hospital of PLA were accessed by both OCM of Germany BeiLang Dialog + hemodialysis machine and single-pool urea kinetic model.The Kt/V calculation value was also checked by the urea dynamics - room variable volume model.The results generated at different time point were compared,including at the real time,before and after appearing through blood by On - line Clearance Monitoring,and correlation analysis of Kt/V value was performed.ResultsThere was no significant difference on evaluating the level of Kt/v by these two methods.The relationship between the results of the two methods was significantly positive related.(r=0.539,P < 0.01).ConclusionsDuring hemodialysis,the OCM option provides an accurate tool for continuous on-line monitoring of urea clearance.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo prognóstico Idioma: Chinês Revista: Journal of Chinese Physician Ano de publicação: 2012 Tipo de documento: Artigo