Analysis on hospital medical equipment electrical safety testing / 中国医学装备
China Medical Equipment
;
(12): 25-26,27, 2015.
Artigo
em Chinês
| WPRIM
| ID: wpr-601385
ABSTRACT
Objective:
To study the electrical safety of multi-parameter monitor and defibrillation monitor, and guarantee the safe and effective operation by through the medical equipment electrical safety performance in our hospital.Methods:
To test 101 different models of multi-parameter monitor and defibrillation monitor electrical safety.Results:
The one-time pass rate of 96%, and the two-time rate was 98%, also one multi-parameter monitor and defibrillation monitor hadn’t meet electrical safety performance after repairing.Conclusion:
The equipments that don’t meet electrical safety performance prohibit using for clinical purposes; To ensure the safety of medical equipment and life safety of medical workers and patients by through developing the testing of the medical equipment electrical safety performance in our hospital.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Idioma:
Chinês
Revista:
China Medical Equipment
Ano de publicação:
2015
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS