ABSTRACT
Objective To analyze the intraoperative and postoperative common complications of Enterprise stent-assisted embolization of intracranial aneurysms and the causes and preventive measures. Methods One hundred forty-three patients with intracranial aneurysm treated with Enterprise stent-assisted embolization at the Department of Neurosurgery,Yijishan Hospital,the First Hospital Affiliated to Wannan Medical College from January 2012 to March 2014 were analyzed retrospectively. The common intraoperative and postoperative complications and its possible causes,as well as the appropriate management were analyzed,and the prognoses were observed. Results A total 143 patients(205 aneurysms)with intracranial aneurysm were enrolled,included 43 with unruptured aneurysm,12 with recurrent aneurysm,and 88 with ruptured aneurysm. A total of 170 Enterprise stents were used. Twenty-two patients (15. 4%)had complications. Among them,2 had intraoperative aneurysm rupture,and they recovered well and discharged after active treatment. Thirteen patients had acute thrombosis,11 of the patients completely restored blood flow immediately after tirofiban and/or urokinase,microcatheter and guidewire-contact thrombolysis. The thrombolysis failed in 1 patient,and the blood flow was slow in 1 patient. Six patients had different degrees of cerebral infarction after procedure,and 1 died (peroperative Hunt-Hess grade Ⅳ). Three patients had vasospasm and they were improved after reducing blood vessel wall irritation and papaverine infusion. The introperative stent guidewire was broken and the stent in place was difficult in 1 case. The last coil packed difficultly during the procedure,and it protruded into the parent artery in 1 case. Two patients had non-aneurysmal hemorrhage after procedure. After conservative treatment,one left unilateral limb muscle strength decline and the other was stable after craniotomy,but leaving aphasia and hemiplegia. Conclusion When using the Enterprise stent-assisted embolization for complex aneurysms,grasping the indications strictly,strengthening the perioperative management and improving the operative skills may reduce or avoid the occurrence of complications.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Cerebrovascular Diseases Ano de publicação: 2015 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Cerebrovascular Diseases Ano de publicação: 2015 Tipo de documento: Artigo