Your browser doesn't support javascript.
loading
The Effect of the Position Change of the Upper Limbs to the Setup Accuracy in the Practice of Immobilizing the Thoracic Cancer Patients by Using Vacuum Bags and Body Frames / 中国医学物理学杂志
Chinese Journal of Medical Physics ; (6): 1472-1473,1484, 2009.
Artigo em Chinês | WPRIM | ID: wpr-605015
ABSTRACT

Objective:

To explore the position change of the setup reference markers on the sternum due to the change of hands grip position.

Methods:

Six hands grip positions were simulated for 8 patients and the set-up reference points marked on the sternum skin were compared while the hands grasped different positions of the hands.

Results:

The setup markers on the sternum were deviated by all five hands grip positions. The maximum deviated distance was 14mm.

Conclusion:

The change of the hands grip position can affect the position of setup reference markers on the sternum of the patients and thus the isocenter of plan may be aligned to the wrong area.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Medical Physics Ano de publicação: 2009 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Medical Physics Ano de publicação: 2009 Tipo de documento: Artigo