OBJECTIVE:
To establish a method for simultaneous residual determination of dichloromethane and ethyl acetate in bisacodyl raw material.METHODS:
Head-space GC was performed on the capillary column of 6% cyanopropyl phenyl-94% di-methyl polysiloxane(DB-624)by temperature programming,the temperature of injector was 220 ℃,detector was flame ionization detector with temperature of 250 ℃,carrier gas was high purity nitrogen with the flow rate of 3.0 ml/min,split ration was 1∶10, headspace heating temperature was 70 ℃,equilibration time was 30 min,volume of headspace vial was 5 ml,and the injection volume was 1 ml.RESULTS:
The linear range was 6-120μg/ml for dichloromethane(r=0.999 9)and 50-1 000μg/ml for ethyl ac-etate(r=0.999 9);the limit of quantitation was 0.2,1.7 μg,limit of detection was 0.06,0.5 μg;RSDs of precision,stability and reproducibility tests were no higher than 3%;recoveries were 100.30%-102.00%(RSD=0.63%,n=9) and 100.10 %-101.30%(RSD=0.44%,n=9).CONCLUSIONS:
The method is simple and accurate,and can be used for the simultaneous residual deter-mination of dichloromethane and ethyl acetate in bisacodyl raw material.