Your browser doesn't support javascript.
loading
Determination of Total Chloride in Oral Rehydration Salt Powder (Ⅲ) by Ion Chromatography / 中国药师
China Pharmacist ; (12): 1689-1691, 2017.
Artigo em Chinês | WPRIM | ID: wpr-607422
ABSTRACT

Objective:

To establish a method for the determination of total chloride in oral rehydration salts powder (Ⅲ) by ion chromatography (IC).

Methods:

The analysis was performed on an IonPac AS 14A (250 mm × 4 mm) column with an IonPac AG 14A (50 mm × 4 mm) guard column and 9mmol·L-1 Na2 CO3 solution as the mobile phase. The flow rate was 1. 0 ml·min-1 with isocratic elution and the sample size was 25 μl. The suppressor was AERS 500(4mm) with 45mA suppression current, and an ECD was used for the detection. The quantity of chloride ion ( Cl-1 ) was calculated by the peak area with an external standard method. Re-sultsThe linear range of Cl-1 was 0. 05-20. 00 mg·L-1(r=0. 9999);the recovery was 98. 89%(RSD=0. 56%, n=6). Conclu-sionThe results demonstrated that the method has the advantages of high sensitivity, simple sample pretreatment, promising accuracy and good reliability, which is suitable for the determination of total chloride in oral rehydration salt powder (Ⅲ) .

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: China Pharmacist Ano de publicação: 2017 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: China Pharmacist Ano de publicação: 2017 Tipo de documento: Artigo