Your browser doesn't support javascript.
loading
Uncertainty Evaluation for the Determination of Salicylic Acid in Human Plasma by UPLC / 医药导报
Herald of Medicine ; (12): 737-740, 2017.
Artigo em Chinês | WPRIM | ID: wpr-620229
ABSTRACT
Objective To evaluate the uncertainty for the determination of salicylic acid in human plasma by ultra performance liquid chromatography(UPLC).Methods All sources of uncertainty in the whole process of salicylic acid determination were analyzed,then the combined and expanded uncertainty were evaluated.Results The expanded uncertainty at concentrations of the lowest limit of quantitation(0.23 μg·mL-1) and a high level(39.43 μg·mL-1) of salicylic acid(P=95%,k=2) was 0.014 and 7.34 μg·mL-1,respectively.Conclusion The uncertainty of salicylic acid determination in human plasma by UPLC was mainly caused by recovery,repeatability and sample preparation at the lowest limit of quantitation and high qulity control concentration.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Herald of Medicine Ano de publicação: 2017 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Herald of Medicine Ano de publicação: 2017 Tipo de documento: Artigo