AlM To introduce a new color pediatric visual acuity chart and its clinical application. METHODS:
The color pediatric visual acuity chart was designed based on principle of visual angle. The optotype on the color chart had graphics. The progression rate of optotype size between 2 lines was 10 10 and 1. 2589. A regular geometric progression of optotype sizes and distribution was employed to arrange 8 lines with 11 optotype on the color chart. The testing distance was 3m. The visual acuity score could be recorded as logarithm of the minimum angle of resolution notation or decimal notation. The reliability of naked distant measurements with this new chart was tested in one eye of 100 children (4 ~ 6 years old) taking the Chinese national standard logarithm visual acuity chart standard.RESULTS:
The color pediatric visual acuity chart and logarithmic chart controls, visual acuity test results that in the two groups had no significant difference (t=1. 2671, P> 0. 05 ). Two sets of vision data existed positive correlation (r= 0. 924, P CONCLUSlONChildren are easier to accept used new color pediatric visual acuity chart to inspect vision. New chart is reliability and apply to children's vision screening.