Your browser doesn't support javascript.
loading
Correlation analysis of fundus examinations with Angio-OCT and FD-OCT / 国际眼科杂志(Guoji Yanke Zazhi)
International Eye Science ; (12): 1543-1547, 2016.
Article em Zh | WPRIM | ID: wpr-637883
Biblioteca responsável: WPRO
ABSTRACT
?AIM:To compare the coherence and difference on the fundus examination made with two kinds of optical coherence tomography ( OCT): Angio-OCT and fourier domain-optical coherence tomography ( FD-OCT) .?METHODS:Using Angio-OCT and FD-OCT to measure the retinal nerve fiber layer ( RNFL ) thickness, optic parameters, and ganglion cell complexes ( GCC ) thickness from 20 subjects respectively.The coherence was tested with Pearson's correlation coefficient, the difference was tested with paired Student t testing.?RESULTS:The total correlation of the RNFL thickness, optic parameters, GCC thickness made with two kinds of OCT was between 0.7-0.8;the RNFL thickness, optic disk area etc.made with the Angio-OCT were lower than those made with FD-OCT except for the GCC thickness.?CONCLUSION: The results made with two kinds of OCT from the same subject has certain coherence, but cannot be compared directly.
Palavras-chave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: International Eye Science Ano de publicação: 2016 Tipo de documento: Article
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: International Eye Science Ano de publicação: 2016 Tipo de documento: Article