The measurement of anterior chamber depth and axial length with the IOLMaster compared with contact ultrasonic axial scan / 国际眼科杂志(Guoji Yanke Zazhi)
International Eye Science ; (12): 921-924, 2007.
Artigo em Chinês | WPRIM | ID: wpr-641669
ABSTRACT
·

AIM:

To compare the measurement of anterior chamber depth (ACD) and axial length (AL) by IOLMaster and contact ultrasonic (US) axial scan (A-scan).·

METHODS:

Measurements of ACD and AL were prospectively obtained in 137 eyes of 121 subjects with the IOLMaster compared with measurements with the US.·

RESULTS:

There was an excellent correlation between IOLMaster and US measurements for the ACD (r=0.823;P<0.001) and AL (r=0.996;P<0.001). The mean values of the parameters measured by IOLMaster and US were,respectively, as follows ACD, 2.94±0.49mm, 2.58±0.51mm;AL, 24.37±3.04mm, 23.81±2.83mm. The mean differences of ACD and AL values between IOLMaster and US measurements were 0.36 ±0.30mm, 0.56 ±0.34 mm respectively, and they proved to be statistically significant (P<0.001), With the 95%limits of agreement (LoA) from -0.08mm to +0.38mm for ACD and from -0.09mm to +0.69mm for AL.·

CONCLUSION:

As noncontact biometry, IOLMaster provides accurate values. A high degree of agreement between US and IOLMaster was noted. It not only has the advantage of performing noncontact examinations, but also produces various additional data simultaneously and may thus obviate the need for multiple examinations. Further studies are needed to assess the interchangeability of measurements in clinical practice.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: International Eye Science Ano de publicação: 2007 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: International Eye Science Ano de publicação: 2007 Tipo de documento: Artigo