Your browser doesn't support javascript.
loading
Central corneal thickness using EX500 excimer laser workstation / 国际眼科杂志(Guoji Yanke Zazhi)
International Eye Science ; (12): 1828-1830, 2014.
Artigo em Chinês | WPRIM | ID: wpr-642037
ABSTRACT

AIM:

To evaluate the accuracy of central conreal thickness ( CCT ) using EX500 Excimer Laser workstation (EX500) in laser in situ keratomileusis (LASIK) patients.

METHODS:

The CCT of 120 eyes (63 patients) who had LASIK between January 2013 and June 2013 were measured by A- scan and EX500. Three groups were classified >550μm, 500 ~550μm,

RESULTS:

The average preoperative CCT value was 527. 9±34. 3μm measured by A-scan, 528. 5±34. 6μm measured by EX500. There was no significant difference between these two measurements (t=1. 736, P=0. 085). In group which CCT >550μm, the average preoperative CCT value was 571. 4±17. 3μm measured by A-scan, 572.7±15. 7μm measured by EX500. There was no significant difference between these two measurements (t=1. 857, P=0. 072). In group which CCT 500 ~ 550μm, the average preoperative CCT value was 523. 4±13. 1μm measured by A-scan, 524. 2±12. 4μm measured by EX500. There was no significant difference between these two measurements ( t=1. 934, P = 0. 058 ). In group which CCT

CONCLUSION:

There is no significant difference between preoperative CCT value measured by A-scan and EX500. After corneal flap lifting and keratomileusis, the CCT value measured by EX500 is smaller than measured by A-scan.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: International Eye Science Ano de publicação: 2014 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: International Eye Science Ano de publicação: 2014 Tipo de documento: Artigo