Excessive Wear of a Polyethylene Insert Caused by IncompleteSeating of a Metal Wire Retaining Ring: A Case Report / 대한정형외과학회잡지
The Journal of the Korean Orthopaedic Association ; : 535-538, 2003.
Artigo em Coreano | WPRIM | ID: wpr-652249
ABSTRACT
We report a case of excessive wear of a polyethylene insert after the incomplete seating of the acetabular metal wire retaining ring in the Omnifit modular noncemented acetabular component (Osteonics Corporation, Allendale, NJ, U.S.A.). Seventy-five months after a primary total hip arthroplasty, metal fragments generated by metal wire breakage in the incompletely seated ring induced excessive abrasive third-body wear at the inner articulating surface and the liner-backing interface. Marked periprosthetic osteolytic lesions, which necessitated revision surgery, were observed. This case suggests that when this type of locking mechanism is used, an intraoperative radiograph should be taken to confirm the proper seating of the retaining ring. When incomplete seating is noticed postoperatively, the patient should be followed up carefully, particularly when the wire is broken.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Artroplastia de Quadril / Polietileno / Acetábulo Limite: Humanos Idioma: Coreano Revista: The Journal of the Korean Orthopaedic Association Ano de publicação: 2003 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Artroplastia de Quadril / Polietileno / Acetábulo Limite: Humanos Idioma: Coreano Revista: The Journal of the Korean Orthopaedic Association Ano de publicação: 2003 Tipo de documento: Artigo