Detection of Residual Brain Arteriovenous Malformations after Radiosurgery: Diagnostic Accuracy of Contrast-Enhanced Three-Dimensional Time of Flight MR Angiography at 3.0 Tesla
Korean Journal of Radiology
;
: 333-339, 2009.
Artigo
em Inglês
| WPRIM
| ID: wpr-65294
ABSTRACT
OBJECTIVE:
Although three-dimensional time-of-flight magnetic resonance angiography (3D TOF-MRA) is used frequently as a follow-up tool to assess the response of arteriovenous malformations (AVMs) after radiosurgery, the diagnostic accuracy of 3D TOF-MRA is not well known. We evaluated the diagnostic accuracy of contrast-enhanced 3D TOF-MRA at 3.0 Tesla for the detection of residual AVMs. MATERIALS ANDMETHODS:
This study included 32 AVMs from 32 patients who had been treated with radiosurgery (males/females 21/11; average patient age, 33.1 years). The time interval between radiosurgery and MRA was an average of 35.3 months (range, 12-88 months). Three-dimensional TOF-MRA was obtained at a magnetic field strength of 3.0 Tesla after infusion of contrast media, with a measured voxel size of 0.40 x 0.80 x 1.4 (0.45) mm3 and a reconstructed voxel size of 0.27 x 0.27 x 0.70 (0.05) mm3 after zero-filling. X-ray angiography was performed as the reference of standard within six months after MRA (an average of two months). To determine the presence of a residual AVM, the source images of 3D TOF-MRA were independently reviewed, focusing on the presence of abnormally hyperintense fine tangled or tubular structures with continuity as seen on consecutive slices by two observers blinded to the X-ray angiography results.RESULTS:
A residual AVM was identified in 10 patients (10 of 32, 31%) on X-ray angiography. The inter-observer agreement for MRA was excellent (kappa= 0.813). For the detection of a residual AVM after radiosurgery as determined by observer 1 and observer 2, the source images of MRA had an overall sensitivity of 100%/90% (10 of 10, 9 of 10), specificity of 68%/68% (15 of 22, 15 of 22), positive predictive value of 59%/56% (10 of 17, 9 of 16), negative predictive value of 100%/94% (15 of 15, 15 of 16) and diagnostic accuracy of 78%/75% (25 of 32, 24 of 32), respectively.CONCLUSION:
The sensitivity of contrast-enhanced 3D TOF-MRA at 3.0 Tesla is high but the specificity is not sufficient for the detection of a residual AVM after radiosurgery.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Malformações Arteriovenosas Intracranianas
/
Sensibilidade e Especificidade
/
Radiocirurgia
/
Angiografia por Ressonância Magnética
/
Meios de Contraste
Tipo de estudo:
Estudo diagnóstico
/
Estudo prognóstico
Limite:
Adulto
/
Feminino
/
Humanos
/
Masculino
Idioma:
Inglês
Revista:
Korean Journal of Radiology
Ano de publicação:
2009
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS