Comparison of Ocular Biometry Using Low-Coherence Reflectometry with Other Devices for Intraocular Lens Power Calculation
Journal of the Korean Ophthalmological Society
;
: 1558-1565, 2015.
Artigo
em Coreano
| WPRIM
| ID: wpr-65426
ABSTRACT
PURPOSE:
To compare axial length (AL) and keratometry (K) using optical low-coherence reflectometry (OLCR, Lenstar LS900(R), Haag-Streit, Bern, Switzerland) with current ocular biometry devices and evaluate the accuracy of intraocular lens (IOL) power calculation.METHODS:
In this prospective, comparative observational study of eyes with cataracts, AL and K were measured using an OLCR device (Lenstar LS900(R), Haag-Streit), partial coherence interferometry (PCI, IOL Master(R), Carl Zeiss, Jena, Germany), A-scan (Eyecubed) and automated keratometry (KR-7100, Topcon, Tokyo, Japan). IOL power calculation was performed using the Sanders-Retzlaff-Kraff (SRK/T) formula. The IOL prediction error (PE) was calculated by subtracting the predicted IOL power from the postoperative (PO) IOL power (PO 4 weeks, PO 12 weeks).RESULTS:
A total of 50 eyes of 39 patients with cataracts (mean age 67.12 +/- 8.51 years) were evaluated in this study. AL and K were not significantly different between the OLCR device and other devices (analysis of variance [ANOVA], p = 0.946, 0.062, respectively). The mean PE in IOL power calculation was -0.22 +/- 0.50D with the OLCR device, 0.08 +/- 0.45D with the PCI device and -0.01 +/- 0.48D with A-scan and automated keratometry (ANOVA, p = 0.006). The highest percentage of eyes with PE smaller than +/- 0.5D was IOL Master(R) followed by Eyecubed and then Lenstar LS900(R). The mean absolute PE was not statistically significant among the 3 devices (ANOVA, p = 0.684).CONCLUSIONS:
Ocular biometry measurements were comparable between the OLCR device and the PCI ultrasound device. However, the IOL power prediction showed significant differences among the 3 devices. Therefore, the differences in application of these devices should be considered.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Catarata
/
Estudos Prospectivos
/
Biometria
/
Ultrassonografia
/
Estudo Observacional
/
Ácido Glicirretínico
/
Interferometria
/
Lentes Intraoculares
Tipo de estudo:
Estudo diagnóstico
/
Estudo observacional
/
Fatores de risco
Limite:
Humanos
Idioma:
Coreano
Revista:
Journal of the Korean Ophthalmological Society
Ano de publicação:
2015
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS