Your browser doesn't support javascript.
loading
Test spatial resolution and average glandular dose of full field digital mammography system / 医疗卫生装备
Chinese Medical Equipment Journal ; (6): 73-76, 2017.
Artigo em Chinês | WPRIM | ID: wpr-659654
ABSTRACT
Objective To test the spatial resolution and average glandular dose (AGD) of full field digital mammography system to obtain high-resolution and contrast breast X-ray images with the lowest dose.Methods The M12 breast X-ray performance test phantom was placed on the breast support center position,then line-pair card was put on the M12 phantom and fixed groove,and exposure was carried out with auto exposure control,28 kV tube voltage and common anode filtration combination so as to detect the spatial resolution.Half-value layer (HVL) was measured by Fluke TNT12000 ray detection tools,2 mm-thickness plate and aluminum sheets with different thicknesses,and then AGD was calculated accordingly.Results The spatial resolutions were all higher than 7 lp/mm and proved to meet the quality standard,which had the vertical spatial resolution being 8 lp/mm,the maximum value of the lateral spatial resolution being 10 lp/mm and the minimum value being 8 lp/mm.AGD was lower than 2 mGy and then proved qualified which was limited within 0.60 and 0.61 mGy.Conclusion The test of spatial resolution and AGD of full digital mammography system can assess the performance and radiation dose,reflect the performance of equipment,ensure image quality and lower radiation dose.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Medical Equipment Journal Ano de publicação: 2017 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Medical Equipment Journal Ano de publicação: 2017 Tipo de documento: Artigo