Your browser doesn't support javascript.
loading
Comparative study on HGB and HCT results by ABL800 blood gas analyzer and BC5800 blood cell analyzer / 医疗卫生装备
Chinese Medical Equipment Journal ; (6): 97-100, 2017.
Artigo em Chinês | WPRIM | ID: wpr-662246
ABSTRACT
Objective To explore the difference between hemoglobin (HGB) and hematocrit (HCT) results by ABL800 blood gas analyzer and BC5800 blood cell analyzer respectively.Methods ABL800 blood gas analyzer and BC5800 blood cell analyzer were used to detect HGB and HCT in 97 samples of heparin-anticoagulated arterial blood and EDTA-3K anticoagulated versus blood from July 17 to 21,2016 as well as 63 samples of heparin-anticoagulated arterial blood from August 4 to 9,2016.Results Almost all the results by ABL800 blood gas analyzer were higher than those by BC5800 blood cell analyzer.Conclusion There're significant differences between ABL800 blood gas analyzer and traditional way when used to detect HGB and HCT.Blood cell analyzer is the preferred choice to detect HGB and HCT,and the results by the blood gas analyzer can be used for references.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Medical Equipment Journal Ano de publicação: 2017 Tipo de documento: Artigo

Similares