Your browser doesn't support javascript.
loading
Voltammetric quantitation of nitazoxanide by glassy carbon electrode / 药物分析学报
Journal of Pharmaceutical Analysis ; (6): 452-455, 2013.
Artigo em Chinês | WPRIM | ID: wpr-672159
ABSTRACT
The present study reports voltammetric reduction of nitazoxanide in Britton-Robinson (B-R) buffer by cyclic and square-wave voltammetry at glassy carbon electrode. A versatile fully validated voltammetric method for quantitative determination of nitazoxanide in pharmaceutical formulation has been proposed. A squrewave peak current was linear over the nitazoxanide concentration in the range of 20-140 mg/mL. The limit of detection (LOD) and limit of quantification (LOQ) was calculated to be 5.23μg/mL and 17.45μg/mL, respectively.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Pharmaceutical Analysis Ano de publicação: 2013 Tipo de documento: Artigo