Your browser doesn't support javascript.
loading
Evaluation of cystatin C with two kinds of detection systems / 中华检验医学杂志
Chinese Journal of Laboratory Medicine ; (12)2003.
Artigo em Chinês | WPRIM | ID: wpr-685658
ABSTRACT
Objective To evaluate the performance of Cys C results among two detection system.Methods The particle-enhanced immunonephrometic assay was used in Dade Behring BNII. Immunoturbic assay was used in Hitachi 7170 to evaluate the JING' YUAN reagents.We compared the precison,linearity,interference,correlation,and calibrators agreement with Dade Behring BNII.Results The total CV of the samples that contain 0.6-5.0 mg/L was less than 10%.The Dade Behring and JING'YUAN method showed good linearity.Haemoglobin(10 g/L),Bilirubin(300 mg/L), Vitamin C(5 g/L)in the tested sample had no significant interference in the assay(interference 0.05) between JING' YUAN and Dade Behring reagents.Values were slightly lower than that from the Dade Behring BNII method,the mean bias was-0.16.The bias range was 1.1%-23% between JING'YUAN and Dade Behring for one sample.Conclusions The precision,linearity and interference test were suitable for routine Cys C measurement on automated biochemistry analyzer,but results has bias.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: Chinese Journal of Laboratory Medicine Ano de publicação: 2003 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: Chinese Journal of Laboratory Medicine Ano de publicação: 2003 Tipo de documento: Artigo