Performance verification of diluted thrombin time assay for detecting Dabigatran level in plasma / 临床检验杂志
Article em Zh | WPRIM | ID: wpr-694790
Biblioteca responsável: WPRO
ABSTRACT
Objective To evaluate the performance of diluted thrombin time (dTT) assay for detecting Dabigatran levels and observe whether this assay may meet the requirements of clinical laboratory.Methods According to EP15-A2,EP6-A,EP7-A and C-24 documents of the Clinical and Laboratory Standards Institute (CLSI),the precision,trueness,analytical measurement range,carryover rate and anti-biological interference of dTT assay were evaluated and the stability of specimen for dTT assay was observed.Results Both the within-day and between-day coefficient of variation (CV) of dTT assay for detecting Dabigatran levels were consistent with manufacturer's stated CV.Compared with target values of Dabigatran,the relative bias of 3 levels of proficiency test materials from College of American Pathologists (CAP) were less than 10%.The results meet linear verification when Dabigatran concentration was between 30.92 and 249.13 ng/mL.The carryover rate was-0.84%.There was no interference for Dabigatran levels by dTT assay for detecting Dabigatran when Hb≤3 g/L,triglyceride≤873 mg/dL,heparin≤2.2 IU/mL and FDP≤29 mg/L.The results of stability showed that plasma specimens for dTT could not be stored at room temperature more than 4 hours,at 4 ℃ more than 4 days,at-20 ℃ exceed 1 month,while at-80℃ the plasma specimens could be stored at least 6 months for dTT assay.Conclusion The precision,trueness,analytical measurement range,carryover rate,anti-biological interference of dTT assay may meet the requirement of clinical laboratory.The stability of the specimen can fulfill the clinical requirements.
Palavras-chave
Texto completo: 1 Índice: WPRIM Tipo de estudo: Guideline Idioma: Zh Revista: Chinese Journal of Clinical Laboratory Science Ano de publicação: 2017 Tipo de documento: Article
Texto completo: 1 Índice: WPRIM Tipo de estudo: Guideline Idioma: Zh Revista: Chinese Journal of Clinical Laboratory Science Ano de publicação: 2017 Tipo de documento: Article