Feasibility of DDD Pacemaker Implantation Under EnSite Systerm Guidance With Micro X-ray / 中国循环杂志
Chinese Circulation Journal ; (12): 178-180, 2018.
Artigo em Chinês | WPRIM | ID: wpr-703838
ABSTRACT

Objective:

To explore the feasibility of DDD pacemaker implantation under EnSite system guidance with micro X-ray.

Methods:

DDD pacemaker implantation under EnSite system guidance with micro X-ray was performed in 5 special patients who needed cardiac pacing while should be avoid X-ray exposure.

Results:

All 5 patients were smoothly finished DDD pacemaker implantation, no wire dislocation or perforation occurred after the operation. The cumulative skin radiation dose was (2.20±0.45) mGy, operative X-ray exposure time was (11.60±2.07) s and the operation time was (58.40±4.04) min.

Conclusion:

DDD pacemaker implantation under EnSite system guidance with micro X-ray was safe and feasible.