Your browser doesn't support javascript.
loading
Uncertainty Evaluation of the Determination of Bilastine in Human Plasma by LC-MS/MS / 中国药师
China Pharmacist ; (12): 1850-1853, 2018.
Artigo em Chinês | WPRIM | ID: wpr-705725
ABSTRACT

Objective:

To establish a method for the determination of bilastine in human plasma, and evaluate the uncertainty by LC-MS/MS.

Methods:

The uncertainty sources were obtained from the whole process of the determination including repeatability, e-quipment error, weighting, solution preparation, calibration fitting and plasma sample handling. The uncertainty and synthesized un-certainty of each component were calculated, and then the expanded uncertainty was obtained.

Results:

The expanded uncertainty for low (15 ng·ml-1), medium (400 ng·ml-1) and high(1 200 ng·ml-1) level of bilastine was 1. 45 ng·ml-1, 28. 72 ng·ml-1 and 74. 61 ng·ml-1, respectively (k=2, P=95% ).

Conclusion:

The uncertainty in the determination of bilastine in human plasma is mainly caused by equipment error, solution preparation, protein precipitation and calibration fitting (especially at low level).

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: China Pharmacist Ano de publicação: 2018 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: China Pharmacist Ano de publicação: 2018 Tipo de documento: Artigo