Determination of Phenol by an Electrochemical Method with Glassy Carbon Electrode / 中国药师
China Pharmacist ; (12): 1878-1880, 2018.
Article em Zh | WPRIM | ID: wpr-705733
Biblioteca responsável: WPRO
ABSTRACT
Objective: To establish an electrochemical method for the determination of phenol. Methods: An electrochemical workstation with three electrodes system was used with glassy carbon electrode as working electrode, Ag/AgCl as reference electrode and Pt as counter electrode. Cyclic voltammetry and differential pulse voltammetry were used for the determination of phenol. Results:Under the condition of 4% Na2SO4as the supporting electrolyte, phenol showed an obvious oxidation peak on the glassy carbon elec-trode. The peak current increased linearly with the concentration of phenol within the range of 0. 8 μg·ml-1-10. 2 μg·ml-1( r=0. 997 5). The lower limit of detection was 0. 20 μg·ml-1. The average recovery was 101. 2% (RSD=2. 2% , n=6). Conclusion:The method is simple and accurate, and can be used for the determination of phenol.