Your browser doesn't support javascript.
loading
Validation and discussion of the time response correction formula for radiation detectors / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 48-51, 2018.
Artigo em Chinês | WPRIM | ID: wpr-708012
ABSTRACT
Objective To validate and discuss the time response correction formula for four types of dosimeters (6150AD6 + 6150AD-b,FH40G + FHZ672E-10,451P ionization chamber and AT1123).Methods The ambient dose equivalent rates shown by survey meters were recorded separately when X-ray emission time was 500,200,100 and 50 ms.The corrected values were obtained by the formula of circuit having a capacitance C and asistance R in series.Results Therewas no correlation between the value measured by AT1123 dosimeter and the time of irradiation.The values by other three kinds of dosimeters obviously varied with the time of irradiation.Conclusions It is not required to make the time response correction for the measured value of ATl123 dosemeter,whereas the values measured by the other three dosimeters could be corrected by the time response correction formula.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiological Medicine and Protection Ano de publicação: 2018 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiological Medicine and Protection Ano de publicação: 2018 Tipo de documento: Artigo